Luminex xPONENT for MAGPIX 4.2 Manuale Utente Pagina 27

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 164
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 26
1. On the Home page, click Shutdown. The Auto Maint tab opens, with System
Shutdown selected.
2. Click Eject.
3. Fill reservoir RA1 with 3/4 of DI water.
4. Fill reservoir RC1 with 3/4 of 10%-20% household bleach solution.
5. Verify that reservoir RD1 is empty.
6. Click Retract.
7. Click Run.
Logging Off and Exiting
To log off and exit xPONENT:
1. Click Logoff at the top of the page.
2. When the Confirm dialog box opens, click OK. This opens the Log In page, with Exit on
the left tab.
3. Click Exit to exit the application.
Using Online Help
English-language help is available at all times while you are using xPONENT. To display
online help for the page or tab in which you are currently working, click the blue “i” icon at the
upper-right of the xPONENT window. This opens a help window with information specific to
that page or tab.
xPONENT for MAGPIX 4.2 Software User Manual
14
Vedere la pagina 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32 ... 163 164

Commenti su questo manuale

Nessun commento