6. Stand the filters upright to air dry.
CAUTION: Filters must be completely dry prior to reinstallation.
7. Reinstall filters. The arrows on the door filters should point to the inside of the door. The
arrow on the XY filter should point up.
8. Reinstall the XY cover and close the access doors.
9. Plug in the power cord and turn on the FLEXMAP 3D instrument.
FIGURE 18.
Ventilation Filter on XY Platform