Luminex 100™ IS User Manual Version 2.3 Lum/nex®

Pagina 2

iv PN 89-00002-00-071 Rev. C Luminex 100 IS User Manual Version 2.3 xMAP TechnologyMaximize/Minimize. . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 3

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 58 PN 89-00002-00-071 Rev. C 2. Select the unprocessed batch that you want to delete. 3. Cl

Pagina 4

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 593. Create a New Batch. The procedure is shown on page 5-53. Click S

Pagina 5

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 60 PN 89-00002-00-071 Rev. C  To open a multi-batch:1. Click Open Multi-Batch. The Open Mu

Pagina 6 - EULA PN: 89-30000-00-070

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 612. Insert the first plate.3. On the Run Batch tab, click Start Plat

Pagina 7 - Contents

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 62 PN 89-00002-00-071 Rev. C 2. Squeeze the barcode reader’s trigger. The beam activates an

Pagina 8 - Basic Concepts 4-1

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 63 To add a Patient List file while creating a batch or multi-batch:

Pagina 9

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 64 PN 89-00002-00-071 Rev. C  To edit a patient list:1. On the File menu, select Edit Pati

Pagina 10

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 653. Select the item that uses the dilution factor that you are setti

Pagina 11 - PN 89-00002-00-071 Rev. C v

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 66 PN 89-00002-00-071 Rev. C  To create a New Advanced Batch:1. Open the Acq. Detail tab.

Pagina 12 - Troubleshooting 7-1

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 676. Click the Bead Set tab. Select the checkboxes next to each desir

Pagina 13 - Glossary A-1

PN 89-00002-00-071 Rev. C vxMAP Technology ContentsDefine Data Export Tab Information. . . . . . . . . . . . . . . .5-41Setting up the Favorites Lis

Pagina 14 - Index Index-1

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 68 PN 89-00002-00-071 Rev. C define commands that apply to one or more wells, one or more r

Pagina 15 - Introduction

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 69Figure 5-38. Plate Layout Tab Command Menu11. Establish preliminar

Pagina 16

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 70 PN 89-00002-00-071 Rev. C Figure 5-40. Plate Layout Tab, Off-Plate Commands12. Insert o

Pagina 17

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 71Managing Assay Lots You can edit standard and control lot informati

Pagina 18

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 72 PN 89-00002-00-071 Rev. C tinue. Once you enter the numbers, the Update Lot Informa-tion

Pagina 19 - Figure 2-2 CE Label

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 734. Enter the expiration date in the Exp. Date field.5. Enter the st

Pagina 20

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 74 PN 89-00002-00-071 Rev. C 4. Click Save. The system updates the lot changes and applies

Pagina 21

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 75 Figure 5-45. Update Lot Information Dialog BoxExport a Lot froman

Pagina 22 - Figure 2-7 Laser Class Label

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 76 PN 89-00002-00-071 Rev. C After responding to the confirmation dialog boxes, regarding s

Pagina 23 - Biological

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 77curve. Replicate samples are averaged to determine the reported qua

Pagina 24 - Blue Indicator Light

vi PN 89-00002-00-071 Rev. C Luminex 100 IS User Manual Version 2.3 xMAP TechnologyDatabase Management Procedures . . . . . . . . . . . . . . . . . .

Pagina 25 - Shipment

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 78 PN 89-00002-00-071 Rev. C Customization Dialog Box Buttons:OK: click to update the graph

Pagina 26

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 79are included in the graph, then Monochrome with Symbols will help d

Pagina 27 - The System

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 80 PN 89-00002-00-071 Rev. C Figure 5-49. Customization Dialog Box—Font Tab• Color Tab—Us

Pagina 28

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 81 Figure 5-50. Customization Dialog Box—Color Tab• Style Tab—use th

Pagina 29

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 82 PN 89-00002-00-071 Rev. C Figure 5-52. Graph Menu ItemsEnable AutomaticAnalysisYou can

Pagina 30

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 832. Click the Auto-Start Analysis checkbox, then click OK. When the

Pagina 31 - Analyzer General

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 84 PN 89-00002-00-071 Rev. C Invalidating standards greatly affects the curve fit and subse

Pagina 32 - Instrument General

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 853. The appropriate Standard or Control dialog box opens. To invalid

Pagina 33 - Electronics

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 86 PN 89-00002-00-071 Rev. C Figure 5-56. Choose Lot Dialog Box—Standard Only3. To create

Pagina 34 - Fluidics

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 87Maintenance Report - provides a history of all maintenance operatio

Pagina 35

PN 89-00002-00-071 Rev. C viixMAP Technology ContentsPressurization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 36 - Figure 3-4 Air Intake Filter

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 88 PN 89-00002-00-071 Rev. C Figure 5-58. Batch Selection Dialog Box3. Enter the informati

Pagina 37 - Connectors

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 89Export Batch Data  To export batch data:1. On the File menu, , cli

Pagina 38 - Reagents

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 90 PN 89-00002-00-071 Rev. C 5. At the Microsoft Windows Print window, select your printer,

Pagina 39 - Basic Concepts

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 91 To back up the database:1. On the Tools menu, click Database Back

Pagina 40

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 92 PN 89-00002-00-071 Rev. C 3. From the Restore Database From dialog box, select a databas

Pagina 41

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 93Drain the Analyzer When draining, you do not need to supply solutio

Pagina 42

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 94 PN 89-00002-00-071 Rev. C  To drain the system:1. On the Maintenance tab, click Eject/R

Pagina 43

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 95 Figure 5-62. Cleanup Utility Dialog Box To perform a disk cleanu

Pagina 44

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 96 PN 89-00002-00-071 Rev. C  To delete the batch folders:1. In the Cleanup Utility dialog

Pagina 45

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 97Perform a SoakCommandSoak the sample probe to prevent the sheath fl

Pagina 46

viii PN 89-00002-00-071 Rev. C Luminex 100 IS User Manual Version 2.3 xMAP TechnologyArchive “My Sessions” folder. . . . . . . . . . . . . . . . . .

Pagina 47

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 98 PN 89-00002-00-071 Rev. C

Pagina 48 - 3. Command List

PN 89-00002-00-071 Rev. C 6 - 1Maintenance and CleaningTo ensure accurate test results, properly clean and maintain the Luminex 100 IS system. Read

Pagina 49

Luminex 100 IS User Manual Version 2.3 xMAP Technology6 - 2 PN 89-00002-00-071 Rev. C 5. Click Wash. In the confirmation dialog box, click on the dro

Pagina 50

xMAP Technology Maintenance and CleaningPN 89-00002-00-071 Rev. C 6 - 3Refill the SheathFluid Container To refill the sheath fluid container:1. Rel

Pagina 51

Luminex 100 IS User Manual Version 2.3 xMAP Technology6 - 4 PN 89-00002-00-071 Rev. C Flush the System Run 3 backflushes, 3 drains, 2 alcohol flushes

Pagina 52

xMAP Technology Maintenance and CleaningPN 89-00002-00-071 Rev. C 6 - 53. Open both doors of the analyzer and clean all accessible surfaces with det

Pagina 53

Luminex 100 IS User Manual Version 2.3 xMAP Technology6 - 6 PN 89-00002-00-071 Rev. C Figure 6-2 Grasping the Tubing4. Remove the filter with one ha

Pagina 54 - Acquisition Detail

xMAP Technology Maintenance and CleaningPN 89-00002-00-071 Rev. C 6 - 7 Figure 6-3 Removing the Screen3. Replace the filter and replace the screen.

Pagina 55 - 4. Acquisition Detail Toolbar

Luminex 100 IS User Manual Version 2.3 xMAP Technology6 - 8 PN 89-00002-00-071 Rev. C Syringe Seal  To replace the syringe plunger seal:1. Disconnec

Pagina 56 - 5. Cancel Command

xMAP Technology Maintenance and CleaningPN 89-00002-00-071 Rev. C 6 - 911. Hand-tighten the screw on the syringe arm.12. Plug in the power cord and

Pagina 57

PN 89-00002-00-071 Rev. C 1 - 1IntroductionAbout This Manual This manual provides you with information to understand and use the Luminex® 100™ IS sy

Pagina 58

Luminex 100 IS User Manual Version 2.3 xMAP Technology6 - 10 PN 89-00002-00-071 Rev. C AnnuallySheath Filter  To change the Luminex 100 analyzer she

Pagina 59

xMAP Technology Maintenance and CleaningPN 89-00002-00-071 Rev. C 6 - 11As requiredFuses To replace the fuses in either the Luminex 100 analyzer or

Pagina 60

Luminex 100 IS User Manual Version 2.3 xMAP Technology6 - 12 PN 89-00002-00-071 Rev. C

Pagina 61 - Disconnected

Luminex 100 IS Maintenance Log Month(s): ________________________________________Year:_______________Maintenance LogProcedure

Pagina 63

PN 89-00002-00-071 Rev. C 7 - 1 TroubleshootingTroubleshooting the Luminex 100 IS SystemTroubleshooting procedures help users isolate, identify, and

Pagina 64

Luminex 100 IS User Manual Version 2.3 xMAP Technology7 - 2 PN 89-00002-00-071 Rev. C 7:00 a.m. and 7:00 p.m. Central Time, Monday through Friday. Inq

Pagina 65

xMAP Technology TroubleshootingPN 89-00002-00-071 Rev. C 7 - 3Pressurization Normal air and sheath pressure readings vary between 6-9 psi while the c

Pagina 66

Luminex 100 IS User Manual Version 2.3 xMAP Technology7 - 4 PN 89-00002-00-071 Rev. C Fluid Leaks Fluid leaks can result in poor pressurization and fa

Pagina 67

xMAP Technology TroubleshootingPN 89-00002-00-071 Rev. C 7 - 5Sample Probe Problems with the sample probe can lead to fluid leaks and pressurization

Pagina 68

Luminex 100 IS User Manual Version 2.3 xMAP Technology1 - 2 PN 89-00002-00-071 Rev. C Technical Support Luminex Technical Support representatives are

Pagina 69

Luminex 100 IS User Manual Version 2.3 xMAP Technology7 - 6 PN 89-00002-00-071 Rev. C Sample arm stuck in the down position.The sample probe height is

Pagina 70

xMAP Technology TroubleshootingPN 89-00002-00-071 Rev. C 7 - 7Calibration and Control ProblemsThe analyzer is misaligned with the XYP instrument.Repo

Pagina 71 - Averaging

Luminex 100 IS User Manual Version 2.3 xMAP Technology7 - 8 PN 89-00002-00-071 Rev. C There is a partial clog in the system.Clean the sample probe. Se

Pagina 72

xMAP Technology TroubleshootingPN 89-00002-00-071 Rev. C 7 - 9Acquisition ProblemsNot enough control microspheres added to the well.Make sure that yo

Pagina 73

Luminex 100 IS User Manual Version 2.3 xMAP Technology7 - 10 PN 89-00002-00-071 Rev. C Bead Detail IrregularitiesUse these tools to assist in diagnosi

Pagina 74

xMAP Technology TroubleshootingPN 89-00002-00-071 Rev. C 7 - 11• error messagesReview reports routinely to detect trends.Use system xMAP control micr

Pagina 75

Luminex 100 IS User Manual Version 2.3 xMAP Technology7 - 12 PN 89-00002-00-071 Rev. C Symptom Possible problem SolutionCalibration microspheres clas

Pagina 76

xMAP Technology TroubleshootingPN 89-00002-00-071 Rev. C 7 - 13Error States The following error states display on the Status Bar on the Run Batch tab

Pagina 77

Luminex 100 IS User Manual Version 2.3 xMAP Technology7 - 14 PN 89-00002-00-071 Rev. C Error Message Possible problem SolutionUnknown Diagnostic Error

Pagina 78

xMAP Technology TroubleshootingPN 89-00002-00-071 Rev. C 7 - 15Sample Error MessagesSample error messages occur when there is a problem with sample a

Pagina 79

PN 89-00002-00-071 Rev. C 2 - 1 SafetySymbols Please become familiar with the information in this chapter before using the equipment. Do not perform

Pagina 80

Luminex 100 IS User Manual Version 2.3 xMAP Technology7 - 16 PN 89-00002-00-071 Rev. C Verify that the system is functioning properly by running the s

Pagina 81 - Tab Information

xMAP Technology TroubleshootingPN 89-00002-00-071 Rev. C 7 - 17Sample probe height is set incorrectly.Adjust the sample probe height.There is a clog

Pagina 82 - Information Tab

Luminex 100 IS User Manual Version 2.3 xMAP Technology7 - 18 PN 89-00002-00-071 Rev. C Luminex SD ProblemsIf the empty sheath fluid container is not r

Pagina 83 - Define Data Export

xMAP Technology TroubleshootingPN 89-00002-00-071 Rev. C 7 - 19Verification  To verify your system from the Maintenance tab:1. Vortex xMAP reagent c

Pagina 84 - Setting up the Favorites

Luminex 100 IS User Manual Version 2.3 xMAP Technology7 - 20 PN 89-00002-00-071 Rev. C 15. Click OK and wait until the Wash completes. The Device Stat

Pagina 85 - Startup Procedures

PN 89-00002-00-071 Rev. C 8 - 1Product NumbersHardware8Product Description Customer NumberRear Air Filter CN-0001-01Bottom Air Filter CN-0002-01Air

Pagina 86

Luminex 100 IS User Manual Version 2.3 xMAP Technology8 - 2 PN 89-00002-00-071 Rev. C Luminex 100 IS Documentation Version 2.3 (Manuals on CD)Luminex

Pagina 87 - Temperature

xMAP Technology Product NumbersPN 89-00002-00-071 Rev. C 8 - 3SoftwarexMAP ReagentsTrainingSyringe Seal (Qty 4) CN-0014-01Cable, USB CN-0018-01Waste

Pagina 88

Luminex 100 IS User Manual Version 2.3 xMAP Technology8 - 4 PN 89-00002-00-071 Rev. C

Pagina 89 - Calibrators

PN 89-00002-00-071 Rev. C A - 1Glossaryagglutination The coalescing of small particles that are suspended in solution; these larger masses are then

Pagina 90

Luminex 100 IS User Manual Version 2.3 xMAP Technology2 - 2 PN 89-00002-00-071 Rev. C Warnings and Notes Informational notes and warnings appear in t

Pagina 91 - Controls

A - 2 PN 89-00002-00-071 Rev. C Luminex 100 IS User Manual Version 2.3 xMAP TechnologyCL1 Refers to dyes embedded in the microsphere. Also see classi

Pagina 92

PN 89-00002-00-071 Rev. C A - 3xMAP Technology Luminex xMAP microsphere setLuminex xMAP microsphere setLuminex multi-analyte microspheres containing

Pagina 93 - CON lots

A - 4 PN 89-00002-00-071 Rev. C Luminex 100 IS User Manual Version 2.3 xMAP Technologysample The mixture of assay components (microspheres, reporter,

Pagina 94 - Templates

PN 89-00002-00-071 Rev. C B - 1Luminex 100 IS System InstallationOverview This appendix provides hardware, firmware, and software installation proce

Pagina 95 -  To create a new batch:

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 2 PN 89-00002-00-071 Rev. C The following sections describe how to connect the Luminex 100

Pagina 96

xMAP Technology Luminex 100 IS System SetupPN 89-00002-00-071 Rev. C B - 3Connect the Luminex 100 analyzer and Luminex XYP to the PCYou must set up

Pagina 97

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 4 PN 89-00002-00-071 Rev. C Figure B-3 Install Alignment Guide7. Ensure that the power sw

Pagina 98

xMAP Technology Luminex 100 IS System SetupPN 89-00002-00-071 Rev. C B - 5Figure B-4 Luminex 100 IS Connections (Luminex 100 analyzer, Luminex XYP

Pagina 99

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 6 PN 89-00002-00-071 Rev. C 12. Connect the color-coded tubing connectors to the Luminex 10

Pagina 100 - Create a Multi

xMAP Technology Luminex 100 IS System SetupPN 89-00002-00-071 Rev. C B - 73. Unscrew the cap from the Cheminert fitting. 4. Insert the sample probe

Pagina 101 - Open a Multi

xMAP Technology SafetyPN 89-00002-00-071 Rev. C 2 - 3Do not perform any maintenance or cleaning of the system’s electrical components, with the exce

Pagina 102 - Process Multiple

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 8 PN 89-00002-00-071 Rev. C 7. Plug the Luminex XYP instrument and the Luminex 100 analyzer

Pagina 103

xMAP Technology Luminex 100 IS System SetupPN 89-00002-00-071 Rev. C B - 92. In a 96-well microtiter plate where overall height is no more than 0.75

Pagina 104

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 10 PN 89-00002-00-071 Rev. C  To install the Luminex XYP instrument reservoir:1. On the Ma

Pagina 105

xMAP Technology Luminex 100 IS System SetupPN 89-00002-00-071 Rev. C B - 113. Place the Luminex SD system near the sheath fluid connection on the le

Pagina 106 - Dilution Factors

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 12 PN 89-00002-00-071 Rev. C • Lower the stainless steel filter end of the sheath fluid lin

Pagina 107 - Advanced Batch

xMAP Technology Luminex 100 IS System SetupPN 89-00002-00-071 Rev. C B - 13Figure B-13 Luminex 100 Analyzer Regulator Adjustment6. From the Home o

Pagina 108

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 14 PN 89-00002-00-071 Rev. C  To install the Luminex XYP instrument heater block:1. On the

Pagina 109 - Color Codes

xMAP Technology Luminex 100 IS 2.3 Software InstallationPN 89-00002-00-071 Rev. C B - 15Luminex 100 Version 1.7 with Windows 98 to Luminex 100 IS Ve

Pagina 110

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 16 PN 89-00002-00-071 Rev. C Figure B-15 Luminex 100 IS 2.3 Startup Splash Screen4. At th

Pagina 111

xMAP Technology Luminex 100 IS 2.3 Software InstallationPN 89-00002-00-071 Rev. C B - 17Figure B-16 Software Information Window3. Run system calib

Pagina 112

© LUMINEX CORPORATION, 2001-2005. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any

Pagina 113

Luminex 100 IS User Manual Version 2.3 xMAP Technology2 - 4 PN 89-00002-00-071 Rev. C The following label appears on the back of the Luminex XYP inst

Pagina 114

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 18 PN 89-00002-00-071 Rev. C 4. The InstallShield Wizard—Welcome dialog box opens. See Figu

Pagina 115 - Unused Template

xMAP Technology Luminex 100 IS 2.3 Software InstallationPN 89-00002-00-071 Rev. C B - 195. Click OK in the Confirmation File Deletion dialog box. If

Pagina 116

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 20 PN 89-00002-00-071 Rev. C 2. Choose the file name and location of the database that you

Pagina 117 - Template

xMAP Technology Luminex 100 IS 2.3 Firmware InstallationPN 89-00002-00-071 Rev. C B - 21Luminex 100 IS 2.3 Firmware InstallationThe Luminex 100 anal

Pagina 118 - Multi-Batches

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 22 PN 89-00002-00-071 Rev. C • Luminex XYP instrument—Continue to use the serial interface

Pagina 119 - Analysis Settings

xMAP Technology Luminex 100 IS 2.3 Firmware InstallationPN 89-00002-00-071 Rev. C B - 23Figure B-19 LXFirmware Update Tool Dialog Box4. Click Upgr

Pagina 120

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 24 PN 89-00002-00-071 Rev. C • If you are upgrading from Luminex 100 IS Version 2.1 or 2.2

Pagina 121

xMAP Technology Luminex 100 IS 2.3 Firmware InstallationPN 89-00002-00-071 Rev. C B - 25Luminex SD System Firmware UpdateIf you are upgrading the fi

Pagina 122

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 26 PN 89-00002-00-071 Rev. C 3. Turn the power off to the SD, XYP, and LX100. 4. Reconnect

Pagina 123

xMAP Technology Installation DrawingPN 89-00002-00-071 Rev. C B - 27Isometric View Right Side ViewLuminex SD19.55 lb. (8.89 kg.)Luminex XYP32.05 lb.

Pagina 124 - Analysis

xMAP Technology SafetyPN 89-00002-00-071 Rev. C 2 - 5You must monitor waste levels manually. Do not allow the waste container to overflow! Empty the

Pagina 125

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyB - 28 PN 89-00002-00-071 Rev. C

Pagina 126

PN 89-00002-00-071 Rev. C C - 1Output.CSVOverview This chapter describes the file specification for the Luminex 100 IS 2.3 OUTPUT.CSV file. Although

Pagina 127

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyC - 2 PN 89-00002-00-071 Rev. C Sample2...SampleMStatNStatN Header: Sample Location, Sample Na

Pagina 128 - Analyses

xMAP TechnologyPN 89-00002-00-071 Rev. C C - 3TemplateName, "value” (Optional) The name of the template used to create the batch. Limited to 3

Pagina 129

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyC - 4 PN 89-00002-00-071 Rev. C Statistics Definitions Statistical calculations are performed f

Pagina 130

xMAP TechnologyPN 89-00002-00-071 Rev. C C - 5Statistics Column DefinitionsThe blocks of statistical data are displayed such that the first row of d

Pagina 131 - Analysis Report

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyC - 6 PN 89-00002-00-071 Rev. C Total Events The number of events that fell with in the defined

Pagina 132 - Procedures

xMAP TechnologyPN 89-00002-00-071 Rev. C C - 7Luminex 100 IS OUTPUT.CSV file with no additional features enabled "Program","Luminex 1

Pagina 133

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyC - 8 PN 89-00002-00-071 Rev. C Luminex 100 IS OUTPUT.CSV file with all additional features ena

Pagina 134 - Recommended Use Schedule

xMAP TechnologyPN 89-00002-00-071 Rev. C C - 9"Samples","8","Min Events","0""Results""DataTyp

Pagina 135

Luminex 100 IS User Manual Version 2.3 xMAP Technology2 - 6 PN 89-00002-00-071 Rev. C All laser apertures are located within the Luminex 100 analyzer

Pagina 136

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyC - 10 PN 89-00002-00-071 Rev. C "3","Std l","4450.92","75&q

Pagina 137 -  To perform a disk cleanup:

xMAP TechnologyPN 89-00002-00-071 Rev. C C - 11"1","Std s","69","75","""2","Std m&q

Pagina 138 - Daily Shutdown

Luminex 100 IS User Manual Version 2.3 xMAP TechnologyC - 12 PN 89-00002-00-071 Rev. C "DataType:","Trimmed StdDev""Location

Pagina 139

PN 89-00002-00-071 Rev. C Index-1Aaccess doors 2-7, 3-9accessoriessee accessory instructions 6-11accuracy/precisionspecifications 3-4add sheath flui

Pagina 140

Index-2 PN 89-00002-00-071 Rev. C Luminex 100 IS User Manual Version 2.3 xMAP Technologycleanaccessible surfaces 6-4Clear Preliminary Off-plate comma

Pagina 141 - Maintenance and Cleaning

PN 89-00002-00-071 Rev. C Index-3xMAP Technology Indexhigh waste volume 6-3Iidlemore than 4 hours 6-1improper system operation 3-3insert off-plate c

Pagina 142 - Routine Tasks

Index-4 PN 89-00002-00-071 Rev. C Luminex 100 IS User Manual Version 2.3 xMAP Technologyphotodiode 4-1photomultiplier 4-1plate 4-1plate holderejectin

Pagina 143 - Note: There is NO warning of

PN 89-00002-00-071 Rev. C Index-5xMAP Technology Indexsheath fluid container 3-11software 3-3, 8-3software overviewLuminex 100 IS software 4-2softwa

Pagina 144 - Clean the Sample Probe

Index-6 PN 89-00002-00-071 Rev. C Luminex 100 IS User Manual Version 2.3 xMAP Technologywarranty 2-2wasteconnector 3-11container 3-12, 6-2high volume

Pagina 145 - Every Six Months

xMAP Technology SafetyPN 89-00002-00-071 Rev. C 2 - 7Barcode Reader Laser This label is attached to the barcode reader.Figure 2-9 Barcode Reader La

Pagina 146 - Air Intake Filter

Luminex 100 IS User Manual Version 2.3 xMAP Technology2 - 8 PN 89-00002-00-071 Rev. C HeatBlue Indicator LightThe blue light above the Luminex 100 an

Pagina 147

xMAP Technology SafetyPN 89-00002-00-071 Rev. C 2 - 9Decontaminating the Luminex 100 Analyzer for Return ShipmentLuminex Technical Support will give

Pagina 148 - 1. Syringe Seal 2. Syringe

Luminex 100 IS User Manual Version 2.3 xMAP Technology2 - 10 PN 89-00002-00-071 Rev. C

Pagina 149 - Ventilation Filter

PN 89-00002-00-071 Rev. C 3 - 1The SystemTheory of Operation Luminex 100 IS technology is based on flow cell fluorometry with Luminex-developed inno

Pagina 150 - Annually

Luminex 100 IS User Manual Version 2.3 xMAP Technology3 - 2 PN 89-00002-00-071 Rev. C of pressurized sheath fluid. A single 20 liter sheath container

Pagina 151 - As required

xMAP Technology The SystemPN 89-00002-00-071 Rev. C 3 - 3xMAP Technology Reagents• Classification calibration microspheres (CAL1)• Reporter calibrat

Pagina 152

Standard Terms and Conditions For Use of ProductBy opening the packaging containing this product ("Product") or by using such Product in any

Pagina 153 - Maintenance Log

Luminex 100 IS User Manual Version 2.3 xMAP Technology3 - 4 PN 89-00002-00-071 Rev. C acquiring sample, running system calibrators, running system con

Pagina 154

xMAP Technology The SystemPN 89-00002-00-071 Rev. C 3 - 555°C (95°F to 131°F) when performing heated assays and using the heater blockLuminex 100 An

Pagina 155 - Troubleshooting

Luminex 100 IS User Manual Version 2.3 xMAP Technology3 - 6 PN 89-00002-00-071 Rev. C Luminex XYP Instrument General• Ambient temperature: 15°C to 30°

Pagina 156 - Problems

xMAP Technology The SystemPN 89-00002-00-071 Rev. C 3 - 7Recommended Additional Equipment UninterruptiblePower Supply(UPS)Luminex highly recommends

Pagina 157

Luminex 100 IS User Manual Version 2.3 xMAP Technology3 - 8 PN 89-00002-00-071 Rev. C Luminex XYPInstrumentVentilation FilterThe XYP instrument ventil

Pagina 158

xMAP Technology The SystemPN 89-00002-00-071 Rev. C 3 - 9The alignment guide directs the sample probe into the Luminex XYP instrument.1. Cheminert F

Pagina 159

Luminex 100 IS User Manual Version 2.3 xMAP Technology3 - 10 PN 89-00002-00-071 Rev. C Air Intake Filter A replaceable air intake filter cleans the ai

Pagina 160

xMAP Technology The SystemPN 89-00002-00-071 Rev. C 3 - 11Sheath Filter The sheath filter removes particles greater than ten microns in diameter fro

Pagina 161 - Control Problems

Luminex 100 IS User Manual Version 2.3 xMAP Technology3 - 12 PN 89-00002-00-071 Rev. C Luminex SheathDelivery SystemFor proper operation, place the L

Pagina 162

PN 89-00002-00-071 Rev. C 4 - 1Basic ConceptsBackground InformationxMAP technology is a versatile system that measures soluble analytes. The Luminex

Pagina 163 - Acquisition

BUYER HEREBY GRANTS TO LUMINEX A NONEXCLUSIVE, WORLDWIDE, UNRE-STRICTED, ROYALTY-FREE, FULLY PAID-UP LICENSE, WITH THE RIGHT TO GRANT AND AUTHORIZE SU

Pagina 164 - Irregularities

Luminex 100 IS User Manual Version 2.3 xMAP Technology4 - 2 PN 89-00002-00-071 Rev. C Photodiodes and a photomultiplier tube receive fluorescent sign

Pagina 165 - Figure 7-2 Normal Histogram

xMAP Technology Basic ConceptsPN 89-00002-00-071 Rev. C 4 - 3System calibrators and controls are provided as part of the system. Calibrate at least

Pagina 166 - Technical Support website

Luminex 100 IS User Manual Version 2.3 xMAP Technology4 - 4 PN 89-00002-00-071 Rev. C

Pagina 167 - Messages

PN 89-00002-00-071 Rev. C 5 - 1Using Luminex 100 IS 2.3 SoftwareThis manual describes how to use the Luminex IS 2.3 software. It includes a glossary

Pagina 168

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 2 PN 89-00002-00-071 Rev. C .Figure 5-1. Luminex IS 2.3 Main WindowThere are five major pa

Pagina 169 - Sample Error

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 3Menu Bar The menu bar contains the following menus: File Menu, Tools

Pagina 170

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 4 PN 89-00002-00-071 Rev. C Toolbar The shortcut buttons on the toolbar can be found in oth

Pagina 171 - ± 3.0. If

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 5Favorites. You can use the Favorites list to create a customized lis

Pagina 172 - Luminex SD

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 6 PN 89-00002-00-071 Rev. C 1. Microtiter plate/reservoir image 2. Temperature and Pressur

Pagina 173

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 7The DD temperature measures the doublet discriminator temperature. D

Pagina 174 - 16. Run four WASH commands

End-User License Agreement (EULA) for Luminex® SoftwareThis Luminex End-User License Agreement (“EULA”) is a legal agreement between you (either an in

Pagina 175 - Product Numbers

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 8 PN 89-00002-00-071 Rev. C Maintenance Tab Use the Maintenance tab to perform maintenance,

Pagina 176

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 9Instructions for maintenance procedures begin on page 5-92. Table 5-

Pagina 177 - Training

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 10 PN 89-00002-00-071 Rev. C The system diagnoses real-time system problems related to flui

Pagina 178

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 11The Detailed Sample Progress section displays the percentage of com

Pagina 179 - Glossary

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 12 PN 89-00002-00-071 Rev. C Figure 5-10. Message LogThe log displays actions in color-cod

Pagina 180

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 13Acquisition Detail tab features:• Batch Name and Description• Batch

Pagina 181

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 14 PN 89-00002-00-071 Rev. C The Batch Name and Description section displays the name of th

Pagina 182

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 15 1. Gate Boundaries 2. Aggregate Beads 3. Numerical Gate Position

Pagina 183 - System Setup

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 16 PN 89-00002-00-071 Rev. C Figure 5-14. Dot Plot Display ExampleFour buttons appear at t

Pagina 184 - 2. PC 5. Luminex SD system

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 17collected events. Post acquisition does not display decaying dot pl

Pagina 185 - XYP to the PC

5. EXPORT RESTRICTIONS. You agree that you will not export or re-export the SOFTWARE PRODUCT to any country, person, entity, or end-user subject to U.

Pagina 186

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 18 PN 89-00002-00-071 Rev. C The Status Bar displays status information as the software pro

Pagina 187

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 19Table 5-4. Status BarCommunication Message DetailsStatus Bar Color

Pagina 188 - Instrument Sample Probe

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 20 PN 89-00002-00-071 Rev. C Secondary WindowsIn addition to the main window, there are oth

Pagina 189

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 21• Microtiter Plate image• Command button group box• Save group box•

Pagina 190

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 22 PN 89-00002-00-071 Rev. C The New Lot button is visible only if the template you are usi

Pagina 191 - Instrument Reservoir

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 23controls, and unknown samples are not always acquired in sequential

Pagina 192 - Calibrate and Verify the

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 24 PN 89-00002-00-071 Rev. C Standards Tab The Standards tab lists all the batch tests with

Pagina 193

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 25Figure 5-17. Analysis Window - Standards Tab OpenWhen a median flu

Pagina 194

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 26 PN 89-00002-00-071 Rev. C Figure 5-18. Non-Competitive AssayExamples of out-of-range sa

Pagina 195 - Instrument Heater Block

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 27Figure 5-19. Competitive AssayThe standards tab displays an Expect

Pagina 196 - Installation

PN 89-00002-00-071 Rev. C iIntroduction 1-1About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1The Luminex

Pagina 197 - Version 2.3

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 28 PN 89-00002-00-071 Rev. C than one background sample, the fluorescence values of the two

Pagina 198 - Verify Successful

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 29Figure 5-20. Analysis Window - Samples Tab OpenReplicateAveragingS

Pagina 199

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 30 PN 89-00002-00-071 Rev. C If the system fails to determine the results for a replicate d

Pagina 200 - 100 Version 1.7

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 31Commands This section describes the different commands used in the

Pagina 201 - 2.1/2.2 to Luminex 100 IS

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 32 PN 89-00002-00-071 Rev. C Connect to Instrument Establishes a connection between the PC

Pagina 202 - 100 IS 2.3

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 33Export Data Exports data to a .csv file in the batch folder. There

Pagina 203 - Firmware

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 34 PN 89-00002-00-071 Rev. C New Advanced Batch Assay developers use New Advanced Batch to

Pagina 204 - Firmware Upgrade

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 35Previous Test This command is found in the Analysis window. Use thi

Pagina 205 - Update Interface

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 36 PN 89-00002-00-071 Rev. C command. However, the sanitize command uses 10% to 20% househo

Pagina 206 - Firmware Update

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 37Start Analysis Opens the Open Batch dialog box, from which you sel

Pagina 207 - Luminex SD System

ii PN 89-00002-00-071 Rev. C Luminex 100 IS User Manual Version 2.3 xMAP TechnologyElectronics . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 208 - First Use

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 38 PN 89-00002-00-071 Rev. C Procedures This section explains how to perform functions in t

Pagina 209 - Installation Drawing

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 39Setting Software Options Set up and customize the system software a

Pagina 210

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 40 PN 89-00002-00-071 Rev. C Auto-Start Analysis. Select this feature to enable the softwar

Pagina 211 - Output.CSV

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 41 Figure 5-22. Options Dialog Box—Company Information TabDefine Dat

Pagina 212 - Field Definitions

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 42 PN 89-00002-00-071 Rev. C Export Location Label Style. Choose one of these options to d

Pagina 213

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 432. Select the command that you want to add to your Favorites list.

Pagina 214 - Statistic Description

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 44 PN 89-00002-00-071 Rev. C When priming, the system draws Luminex xMAP sheath fluid from

Pagina 215 - Definitions

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 45distilled water after Sanitize. Place at least 200 µL in a microtit

Pagina 216 - Column Name Description

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 46 PN 89-00002-00-071 Rev. C  To set the Luminex XYP instrument heater temperature:1. Clic

Pagina 217

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 47Ensure that the Luminex analyzer lasers are warmed up and the probe

Pagina 218

PN 89-00002-00-071 Rev. C iiixMAP Technology ContentsStatus Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-17S

Pagina 219

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 48 PN 89-00002-00-071 Rev. C Figure 5-25. Update CAL Targets Dialog Box10. Repeat step 9

Pagina 220

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 49You must run system controls following calibration. Continue with t

Pagina 221

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 50 PN 89-00002-00-071 Rev. C 4. Ensure that the analyzer is set to draw CON1 and CON2 beads

Pagina 222

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 512. Click Import CAL or Import CON as appropriate. The Open dialog b

Pagina 223

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 52 PN 89-00002-00-071 Rev. C You can group batches together as a multi-batch. Multi-batches

Pagina 224

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 53 To import a template from a diskette or CD:1. Insert the kit’s di

Pagina 225

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 54 PN 89-00002-00-071 Rev. C 5. If you want to insert an Acquire Patient or Skip command, s

Pagina 226

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 55If all patient IDs in the batch are identified, the system appends

Pagina 227

Luminex 100 IS User Manual Version 2.3 xMAP Technology5 - 56 PN 89-00002-00-071 Rev. C 2. Click Clear in the dialog box.3. Click Yes to confirm that

Pagina 228

xMAP Technology Using Luminex 100 IS 2.3 SoftwarePN 89-00002-00-071 Rev. C 5 - 572. Select the desired batch under the My Batches folder and click O