Luminex® xPONENT® 3.0 Upgrade Instructions to xPONENT® 3.1 SP1 For In Vitro Diagnostic Use Lum/nex®
Upgrading Luminex xPONENT Release 3.0 to Release 3.1 SP1 - For In Vitro Diagnostic Use6To install the software:1. Click Install xPONENT to begin inst
7To install the xCONVERT Data Conversion Utility:1. Click Install xPONENT file conversion utility. A File Download – Security Warning dialog box open
Upgrading Luminex xPONENT Release 3.0 to Release 3.1 SP1 - For In Vitro Diagnostic Use8Verifying the FirmwareVerify that the firmware version for the
© Luminex Corporation, 2001-2009. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any
iiiTable of ContentsRequirements... 1F
ivUpgrading Luminex xPONENT Release 3.0 to Release 3.1
1For ease in upgrading Luminex xPONENT software from 3.0 to 3.1 SP1, Luminex supplies an upgrade kit. These instructions describe the requirements and
Upgrading Luminex xPONENT Release 3.0 to Release 3.1 SP1 - For In Vitro Diagnostic Use2Your Luminex system should also have a Luminex XYP with a UL ma
3Upgrade Kit ContentsThese items come with all upgrade kits:• Installation CD for xPONENT 3.1 SP1, including Release Notes. This also includes a serie
Upgrading Luminex xPONENT Release 3.0 to Release 3.1 SP1 - For In Vitro Diagnostic Use4WARNING: While upgrading your Luminex system, take all standard
5Uninstalling Luminex xPONENT 3.0After backing up your data files, uninstall xPONENT 3.0. Perform this step only if you did not get a new PC for your